Flocare Infinity Enteral Feeding Sets (Bottle Set · no port)

Product information

  • Quantity Unit Single
  • Product Code 557360564

$11.20 incl GST

Description

Nutricia Flocare Infinity enteral feeding sets are single‑use administration sets engineered for use exclusively with the Flocare Infinity pump. Each set uses a dedicated pump cassette that seats securely in the pump; a looped silicone segment engages the rotor and requires correct door closure for operation. Sets are intended for continuous or preset‑dose delivery as configured on the pump.

Compatibility and connections

  • Patient end: Female ENFit connector compliant with ISO 80369‑3; supplied with a protective dust cap.
  • Feed end options:
  • Pack Sets: ENPlus cross‑spike for direct connection to compatible enteral nutrition packs.
  • Bottle Sets: Universal connector for wide‑neck (40 mm, PT40) or crown‑cork bottles; includes a drip chamber and bottle hanger.
  • Medication ports (on applicable variants): ENFit‑compatible Y‑port or 3‑way stopcock.

Variants (family)

  • Pack Sets: No port; Y‑port; 3‑way port with drip chamber.
  • Bottle Sets: No port with drip chamber; Y‑port with drip chamber; 3‑way port with drip chamber.
  • Pack Mobile Sets (short length for mobility): No port; Y‑port.

Typical components include the Infinity pump cassette with silicone loop, drip chamber (where applicable), medication port (where applicable), line clamp on some sets, female ENFit patient connector with cap, and bottle hanger on Bottle Sets.

Dimensions and materials

  • Standard tubing length: approximately 2300 mm.
  • Mobile set length: approximately 1580 mm.
  • Internal tubing diameter (Pack and Bottle sets): approximately 2.7 mm.
  • Material: Medical‑grade PVC; latex‑free; DEHP‑free.

Medication port positioning is typically about 40 cm from the patient end on Bottle Sets and standard Pack Sets; on Pack Mobile Sets it is positioned closer to the patient end (approximately 9 cm).

Pump setup and priming

  • Insert the cassette fully into the pump and close the door.
  • Use the pump’s FILL SET function to automatically prime; typical fill takes about one minute and stops near the end of the set. If the line is not completely filled, repeat FILL SET until full.

Use and change‑out

  • After priming, connect to the patient’s ENFit feeding tube.
  • Single‑use only; replace the set every 24 hours.

Handling notes

  • Pack Sets: Designed to pierce the foil seal of compatible packs; do not pre‑pierce with other devices.
  • Bottle Sets: Remove the bottle’s foil seal completely before attaching; do not use the set to pierce foil.
  • Storage: Keep clean and dry; protect from dust, rain, and direct sunlight.

Alarm cues related to the set

  • NO SET: Feeding set not inserted.
  • AIR: Air detected; re‑prime with FILL SET and check insertion/air sensor area.
  • OCC IN/OUT: Possible occlusion; inspect the line for blockage.
  • DOOR: Pump door not correctly closed.