ALGERBRUSH - Type: 2 CHUCK-2
Product information
- Quantity Unit Single
- Product Code None
Sorry. This item is out of stock
Description
The ALGERBRUSH II by Bausch + Lomb is a specialized, battery-powered rotary instrument designed for precise ophthalmic procedures. Its primary function is the effective removal of corneal and scleral rust rings and foreign bodies, making it an essential tool for ophthalmologists, optometrists, ER physicians, and veterinarians.
Key Features
- Compact and Lightweight: With a length of approximately 120 mm, the ALGERBRUSH II is shorter than a ballpoint pen, offering portability and ease of use.
- Low-Torque Motor: Powered by a single AA battery, the device features a low-torque, single-speed motor that ensures gentle operation.
- Rotary Burr Tips: Equipped with carbide or diamond burr tips, it gently debrides corneal tissue, minimizing damage and promoting healing.
- Integrated On/Off Switch: The end nut functions as an on/off switch, with an internal clutch mechanism to prevent excessive pressure and reduce injury risk.
- Interchangeable Burrs: Available in various sizes and grits, the burrs are designed for versatility and precision.
Design and Ergonomics
- Quiet Operation: Ensures patient comfort during use.
- Pen-Like Design: Provides ergonomic control for both right and left-handed users, enhancing precision.
Usage Instructions
- Burr Assembly: Ensure a snug fit before use and operate the device at an angle to avoid corneal penetration.
- Battery Installation: Install with the negative pole next to the motor and remove if storing for over 30 days.
- Sterilization: Only the chuck and burr assembly should be sterilized; the main housing can be cleaned with alcohol wipes.
Maintenance
- Sterilization Methods: Use validated steam cycles and avoid chemical or dry heat methods.
- Cleaning: Employ pH-neutral solutions, sonicate, rinse, and inspect burrs for damage.
- Storage: Store burrs in sterilization pouches and inspect for wear before reuse.
Safety Precautions
- Professional Use: Intended for trained professionals with prescription use only.
- Environmental Conditions: Operate within 15–35°C temperature, 25–90% humidity, and 80–110 kPa pressure.
- Battery Management: Ensure the battery is removed during autoclaving and long-term storage.
Regulatory Information
- CE Marked: Compliant with medical device standards since 1998.
- Distribution: Not available in Canada due to regulatory requirements.
Replacement Parts
Replacement burrs and other components are readily available, ensuring the ALGERBRUSH II remains a reliable and versatile tool in various ophthalmic settings. Its longstanding presence in the medical field underscores its effectiveness and trusted performance worldwide.